Customer Feedback Survey
26,811 Products and Counting!

CardioChek Analyzer - Frequently Asked Questions

CardioChek Analyzer - Frequently Asked Questions

CardioChek Analyzer - Troubleshooting 2

What if my analyzer reads CHECK STRIP FAILED?

  • Inspect the optical glass in the analyzer where the test strip is inserted. Is it clean, dry and free from fingerprints? If not, clean with alcohol wipe and dry completely with gauze or lint-free cloth or tissue.
  • Is the check strip damaged or dirty? If so, clean with soap and water or alcohol wipe, or replace.
  • Has the strip been inserted right side up along the guides?
  • Has the meter been dropped or stored in extreme conditions?
  • Is the analyzer being operated inside, away from direct sun or lamp light?
  • Have you checked the batteries?

What if I get "empty" or unexpected results?

  • Results are above or below measuring range.
  • High temperatures affect the strips. Has the vial of strips been stored in extreme conditions? Has the vial been tightly capped immediately after removing the test strip? (The test strips should not be removed from the vial for long periods of time because the strips are adversely affected by humidity.) Is the comparative laboratory method the same as the CardioChek method? Are you using whole blood? (Ketone is the only test strip that uses whole blood, serum, or Plasma. All other test strips require fingerstick whole blood or venous whole blood collected in either EDTA or heparin tubes.) Is an adequate sample size used? Too much or too little blood may give an inaccurate result. Check your package insert for the correct sample size. (Most single tests require 15 µL; multiple test panels require 25 µL - 40 µL.) A capillary blood collector, capillary tube, or pipet must be used to ensure accurate dosing. Did you apply blood to the test strip with the test strip inserted in the analyzer? If not, the analyzer may not produce correct results. Have you changed the batteries? Does the check strip pass? Has the optical glass in the analyzer (where the strip is inserted) been cleaned properly? Check the back of the strip after testing. The membrane must be a uniform color. If there are any blotches, the test results may be affected. Test again with a fresh, unused test strip.

What if the analyzer reads LOT EXPIRED?

  • What is the expiration date on the vial of strips?
  • If the strips are not past their expiration date, check to make sure the date on the analyzer is set correctly. To check the date:
    1. Press the "Enter" button enter until you see the RUN TEST display.
    2. Press the "Next" button next until USER SET displays.
    3. Press the "Enter" button enter once and then press the "Next" button next until SET DATE displays.
    4. Press the "Enter" button enter three times to display MONTH, DAY and YEAR.
  • To change the date, follow steps 1-4 above, but press the "Next" button next followed by the "Enter" button enter to change the date. (Or you can turn the CardioChek off and turn it on and watch the display until the date is displayed.)

What if the CardioChek PA or CardioChek analyzer will not recognize the MEMo Chip? Am I inserting the MEMo chip correctly?

  • Is the MEMo Chip installed properly? The notched side of the MEMo chip faces up.
  • Was the chip inserted before the analyzer was turned on? If so, press the "Enter" button enter twice so the analyzer can recognize that the MEMo Chip has been inserted.

What if Test results are in decimals or appear to be higher or lower than expected?

  • The UNITS menu needs to be reset from INTL to US or vice versa. To change units:
    1. Press the "Enter" button enter until you see RUN TEST display.
    2. Press the "Next" button next until USER SET displays.
    3. Press the "Enter" button enter once and then press the "Next" button until UNITS displays.
    4. Press the "Enter" button enter to view units (either US or INTL). Press the "Enter" button enter to accept current setting or the "Next" button next followed by the "Enter" button enter to change the units.

How can I collect a blood sample?

  • Fingerstick whole blood may be collected in capillary tubes or collectors. Venous whole blood samples can be collected in EDTA tubes (Lavender tops) or Heparin tubes (Green top).

How much blood is needed for each test strip?

  • Single chemistry tests (Cholesterol, HDL, LDL, Triglycerides, Glucose, and Ketone), require 15 µL of whole blood. Lipid Panel tests require 35-40 µL of whole blood. Chol+Glu test panels require 25-30 µL of whole blood. Creatinine tests require 20 µL of whole blood.

How do I change settings on a CardioChek PA or CardioChek analyzer?

  • To change the settings on a Cardiochek PA or CardioChek analyzer, go to the USER SET prompt and press the "Enter" button . Follow the prompts.

The Control Solution I received does not list the value I need to test the lot of Strips I received. What should I do?

  • The Control Solution was most likely manufactured and distributed prior to the lot of strips that you have. You can call PTS Customer Service Toll-free (U.S.) at (877) 870-5610, and they will provide you with the values that you need. You can also click here to look up your values on this website.

How do I calculate my LDL?

  • CardioChek Lipid Panel Test Strips will automatically calculate LDL values (require CardioChek PA analyzer). If using individual test chemistries with either CardioChek PA or CardioChek analyzers, first test Cholesterol, HDL Cholesterol and Triglycerides using new, unused test strips. Then use the following formula to calculate: LDL Cholesterol = Total Cholesterol - HDL - (Triglycerides/5)

More Troubleshooting Questions!

Browse our complete line of Polymer products

  • QuickMedical
  • Copyright 2009
  • Toll Free: 888-345-4858
  • Local: 425-222-5963
  • Fax: 425-222-6030